Dental Titanium Discs — CAD/CAM Disc Blanks for Milling Centres
Dental titanium CAD/CAM disc blanks in commercially pure Grade 1 / 2 / 3 / 4 to ASTM F67 and ISO 5832-2, and Grade 23 Ti-6Al-4V ELI to ASTM F136 and ISO 5832-3. Open system Ø98 mm with central register step, Zirkonzahn Ø95 mm and Amann Girrbach Ø89 mm with side notch. Vacuum annealed for residual stress relief, CMM verified, EN 10204 3.1 with heat-level batch traceability, manufactured under an ISO 13485:2016-certified authorised contract manufacturing partner. MOQ and schedule are quotation-specific.
Read moreShow less
Ø98 / 95 / 89 mmInterface diameters
CP Gr1-4 / Gr23Grade family
ASTM F67 / F136Material standards
Quotation-specificLead time
ASTM F67 ASTM F136 ISO 5832-2 ISO 5832-3 ISO 13485:2016 EN 10204 3.1
Product
Dental Titanium Disc 98 mm CAD-CAM
The Ø98 mm open-system blank, in the published configuration: the grade, the diameter, the thickness ladder, the processing route and the commercial baseline, stated in full so a laboratory can compare it against a quotation line by line.
Pair of Ø98 mm blanks, machined faceSingle blank, polished rim and machined faceThickness comparison, machined edge profileStacked blanks, consistent diameter and face finish
Parameter
Published Value
Note
Material
GR2 (CP Grade 2) / GR5 ELI / TC4 ELI / GR23 (Ti-6Al-4V ELI)
ASTM F67 for CP, ASTM F136 for the ELI alloy
Dimension
Ø98 mm × TK 10–25 mm
Ø98 mm open-system nominal geometry
Standard
ASTM F136 / ASTM F67
ISO 5832-3 and ISO 5832-2 equivalents
Processing
Water cutting and turn milling
Blanks cut from certified plate, then machined to the interface geometry
Surface
Polished or machined
Specified per order against the machining route
MOQ
30 pieces
Catalogue baseline for this item; the confirmed quantity is set in the quotation
Delivery time
1–3 days ready for shipment
Catalogue baseline from released stock; the confirmed schedule is set in the quotation
Customizable
Dimension, logo, package
OEM/ODM marking, engraving and packing options
Payment terms
70% advance payment, balance before delivery
Standard terms for catalogue orders
Certified by
ISO 9001:2015 quality management, with manufacturing under an ISO 13485:2016 partner
The blank carries material conformity, not device registration — see the note below
Available thicknesses (mm)
Eight published thicknesses on the Ø98 mm blank. The thickness is matched to the machine pocket depth and to the height of the restoration the laboratory intends to mill.
Reading these values correctly: the minimum order quantity, the readiness time and the payment terms above are the catalogue baseline for this item, and the values confirmed for a specific order are the ones stated in the quotation. The blank carries material conformity to ASTM F67 and ASTM F136 under an ISO 13485:2016-certified manufacturing partner and BOZE CNC-Ti's ISO 9001:2015 quality management system; it is a semi-finished material product, not a device registration, and FDA or CE device registration remains the responsibility of the laboratory or manufacturer that mills the final restoration.
Product — Dental Titanium Disc 98 mm CAD-CAM
Dental Titanium Disc 98 mm CAD-CAM — The Ø98 mm Open-System Blank
Our Ø98 mm disc is the open-system CAD/CAM blank made for guided dental milling: the diameter and the locating feature that a standard open-interface chuck actually engages, delivered as a ready-to-clamp blank so the laboratory mills restorations rather than cutting its own discs. It is sourced as certified plate, water-cut and turn-milled to the published interface geometry, then vacuum annealed and dimensionally verified before release.
Premium material — supplied in CP Grade 2 to ASTM F67 and in Grade 23 Ti-6Al-4V ELI to ASTM F136, both with biocompatibility and mechanical properties suited to implant and prosthetic use
Certified quality — manufactured to ASTM F67 and ASTM F136 under an ISO 13485:2016-certified authorised contract manufacturing partner
High durability — a stable, stress-relieved blank with the wear resistance and mechanical stability needed for long-lasting restorations
BOZE Metal is a sourcing and processing partner for titanium disc blanks. We source certified bar and plate, contract the manufacturing to an ISO 13485-certified partner, and control the order basis, geometry, inspection scope and documentation. The following scope clarifications matter for accurate technical and commercial alignment.
🏥
Not a medical device registration
A titanium blank is a semi-finished material product. It carries material conformity to ASTM F67 or ASTM F136, not a device approval. Device-level certification, clinical validation and CE or FDA registration remain the responsibility of the dental laboratory or manufacturer who mills the final restoration.
📜
The ISO 13485 certificate is the partner's
Manufacturing is performed under a contract manufacturing partner that holds ISO 13485:2016 and is authorised to supply BOZE. The certificate is issued to that partner, not to Baoji Boze Metal Products Co., Ltd. The certificate scope and the authorisation letter are available on request.
🏭
Not a disc mill
We do not roll or press titanium disc. Discs are sourced as bar and plate and machined to the interface geometry. Our scope is grade and specification review, interface geometry agreement, machining coordination, inspection, documentation and shipment.
🤝
Brand names identify geometry only
Roland, VHF, Imes-Icore, Yenadent, Zirkonzahn and Amann Girrbach are referenced solely to identify the published chuck geometry each blank is machined to. No affiliation, partnership or endorsement by those manufacturers is claimed or implied.
🦷
Not finished prosthetics
We supply the starting blank. Crowns, abutments, full-arch frameworks, screw-retained prostheses and any other finished restoration are milled from it by the dental laboratory or milling centre that owns the machine.
⏱️
No fixed MOQ or lead time
Minimum order quantity and schedule depend on grade, interface geometry, blank dimensions, quantity, current bar and plate availability and the machining plan. Treat them as quotation-specific values confirmed during engineering review.
Order Basis
Freeze Four Inputs Before Requesting a Quotation
A disc blank is fully defined by four inputs. State all four and the quotation can be produced on the first exchange instead of after three clarification rounds.
Interface and machine
Name the milling system and the exact blank diameter and feature geometry it expects. A Ø98 mm blank will not enter a Ø95 mm or Ø89 mm chuck, and the mismatch is discovered at the machine, not at the purchasing desk.
Grade and specification
State the grade and the governing standard revision — ASTM F67 or ASTM F136 for the material, or the ISO 5832-2 / 5832-3 equivalent. Grade 2 and Grade 23 are not interchangeable in a clinical supply chain even when dimensions match.
Blank geometry
Give the thickness and the machined-part height the laboratory intends to produce, so the blank can be sized against the machine pocket and the planned restoration height.
Documentation scope
Define the mill certification level, the traceability granularity, the inspection characteristics required and the third-party witness arrangement before the order is placed.
Interface Geometry
Match the Blank Diameter and Locating Feature to Your Machine
Three published disc geometries cover the dental CAD/CAM installed base. The diameter and the locating feature are what the chuck actually engages, so both must match. Brand names identify the published geometry only.
Interface
Blank Diameter
Locating Feature
Machines Served
Notes
Open system
Ø98 mm (nominal)
Central register step
Roland, VHF, Imes-Icore, Yenadent and other open-interface chucks
The most common CAD/CAM disc geometry. Chucks clamp on the outer diameter with the step locating the blank axially.
Zirkonzahn system
Ø95 mm
Zirkonzahn-specific blank geometry
Zirkonzahn 5-axis milling units
A dedicated interface, not interchangeable with the Ø98 mm open-system blank.
Amann Girrbach system
Ø89 mm
Side notch locating feature
Amann Girrbach milling centres
Located by a peripheral notch rather than a central step. Not interchangeable with Ø95 mm or Ø98 mm blanks.
The three geometries are not interchangeable. A Ø95 mm blank will not enter a Ø98 mm chuck even though both are described as "titanium discs". If your machine is a hybrid or a retrofit, send the machine model and the current blank part number and the interface is confirmed from those.
Grade Selection
Commercially Pure and ELI Grades for Dental Restorations
Five grades, each with a different strength-to-ductility balance. Values are the typical annealed-condition minima recorded for each grade in the titanium grade library; the mill test report for the order governs acceptance.
Grade / UNS
Rm (min)
Rp0.2 (min)
A5 (min)
Oxygen
Typical Dental Use
Grade 1UNS R50250 · CP Titanium
≥ 240 MPa (35 ksi)
≥ 170 MPa (25 ksi)
≥ 24%
Lowest interstitial of the CP grades
High-ductility frameworks, burners, and full-arch structures where formability dominates
Grade 2UNS R50400 · CP Titanium
≥ 345 MPa (50 ksi)
≥ 275 MPa (40 ksi)
≥ 20%
Typical max 0.13% for mill-stock supply
Single crowns, abutments and general prosthetics — the most widely specified CP grade
Grade 3UNS R50550 · CP Titanium
≥ 450 MPa (65 ksi)
≥ 380 MPa (55 ksi)
≥ 18%
Elevated by design — solid-solution strengthened
Higher-load abutments and copings where CP Grade 2 strength is insufficient
Grade 4UNS R50700 · CP Titanium
≥ 550 MPa (80 ksi)
≥ 485 MPa (70 ksi)
≥ 15%
Highest CP oxygen ceiling — highest CP strength
Load-bearing implant components and surgical hardware; the strongest CP grade
Grade 23 (Ti-6Al-4V ELI)UNS R56401 · α-β ELI alloy
≥ 860 MPa (125 ksi)
≥ 795 MPa (115 ksi)
≥ 10%
≤ 0.13% maximum — extra-low interstitial
Implant abutments, screw-retained frameworks and fatigue-critical components
CP Grade 1-4 are supplied to ASTM F67 and ISO 5832-2. Grade 23 Ti-6Al-4V ELI is supplied to ASTM F136 and ISO 5832-3.
Substituting Grade 5 for a Grade 23 callout fails a medical device audit even when the dimensional results are equivalent, because the MTR must declare the correct UNS number.
Specification Reference
Dental Titanium Disc Blank Specification EAV
Twelve procurement-critical attribute, value and standard references. Use this as a cross-check table when drafting an RFQ; the applicable specification is selected per order.
Attribute
Value / Range
Standard
Blank outside diameter
Ø98 mm open system / Ø95 mm Zirkonzahn / Ø89 mm Amann Girrbach
CAD/CAM interface geometry
Blank thickness
Commonly 10-16 mm; confirmed against machine pocket depth and part height
Machine capability
Material grades
CP Grade 1 / 2 / 3 / 4 and Grade 23 Ti-6Al-4V ELI
ASTM F67, F136, ISO 5832-2, 5832-3
Interstitial control (ELI)
Oxygen ≤ 0.13% maximum on Grade 23; nitrogen, carbon and hydrogen held to the same low-interstitial envelope
ASTM F136 / ISO 5832-3
Grain size control
Fine equiaxed alpha-beta microstructure reported per the accepted grain-size method
ASTM E112
Heat treatment
Vacuum annealing — stress relief at 700-800°C for Grade 23 where specified
Vacuum cycle per order
Hardness
Reported as micro-Vickers or the equivalent scale named on the order
ASTM E18
Dimensional inspection
CMM verification of diameter, thickness, step height and concentricity
Order specification
Surface defect inspection
Dye penetrant inspection where surface-defect acceptance is specified
ASTM E1417
Mill certification
EN 10204 Type 3.1 standard; Type 3.2 with independent witness on request
EN 10204
Batch traceability
Heat identity carried from incoming bar or plate through to the shipped blank
EN 10204 3.1
Manufacturing QMS
Manufactured under an ISO 13485:2016-certified contract manufacturing partner; authorisation available on request
ISO 13485:2016
Chemical Composition Control
The Interstitials That Separate Medical-Grade Titanium from Industrial Titanium
A titanium heat can carry the right UNS number on the paperwork and still fail in the mouth. The interstitial elements decide whether the material is genuinely implant-grade or only nominally so.
Element
Why It Is Controlled
Control Basis
Standard
Oxygen (O)
Primary solid-solution strengthener in titanium and the dominant ductility killer. It raises tensile strength and simultaneously reduces elongation, toughness and as-machined integrity.
≤ 0.13% max on Grade 23 ELI
ASTM F136 / ISO 5832-3
Nitrogen (N)
The most damaging interstitial for ductility. Above specification it embeds at grain boundaries and is a primary contributor to hydrogen-assisted cracking and embrittlement.
Held to the low-interstitial envelope; reported on the MTR
ASTM F136 / ISO 5832-3
Hydrogen (H)
Diffuses into titanium during pickling, heat treatment and extended exposure to hydrogen-bearing environments, promoting delayed hydrogen-assisted cracking.
Vacuum-annealed processing; reported on the MTR
ASTM F136 / ISO 5832-3
Iron (Fe)
A common mill contaminant that forms brittle intermetallic phases, reducing ductility and fatigue life. Fe is the usual difference between a certified implant-grade heat and a general industrial heat.
Reported on the MTR against the specification
ASTM F67 / F136
Carbon (C)
Forms carbides at grain boundaries when elevated, again reducing toughness and fatigue resistance under cyclic loading.
Reported on the MTR against the specification
ASTM F67 / F136
Every heat is reported against the specification on the mill test report. It is that declared result, not the grade name, which governs acceptance — and it is the document your quality system will audit.
Milling Intelligence
Milling Titanium Disc Blanks Without Burning the Edge
Titanium is the most difficult common dental metal to machine. These are the principles that govern a stable setup, stated as relationships rather than absolute numbers, because the correct surface speed depends on your machine, not on a table.
Parameter
Guidance
Boundary
Cutting speed
Reduce relative to steel on the same machine and the same tool material. Titanium generates heat at the cutting edge faster than steel because of its low thermal conductivity and its tendency to work-harden under dwell.
The absolute surface speed is set by machine rigidity, holder length, tool diameter and the blank grade — not by a table value.
Feed per tooth
Keep the feed per tooth stable and relatively high for the tool diameter rather than reducing it. A light, dwelling feed promotes work hardening and accelerates edge breakdown on CP titanium.
Excessive feed on a thin disc risks deflection and step breakout; the acceptable combination is confirmed during the setup review.
Coolant delivery
Flood or through-spindle coolant with sufficient pressure to reach the cutting edge and the full tooth space. Intermittent or starved cooling is the most common cause of titanium edge burn.
Coolant delivery capability varies by machine and by holder; confirm whether through-tool supply is available before the parameter set is committed.
Tool strategy
Sharp, positive-rake carbide or a variable-helix geometry suited to titanium, with a rigid setup. Variable helix spreads engagement along the edge and reduces the heat concentration that causes burn.
Coated tooling selection is a function of the specific alloy, the cut length and the coolant present.
Blank stability
Vacuum annealing before roughing removes residual stress from the machining of the blank itself, so the disc does not move or distort once the laboratory begins cutting abutment geometry.
Residual stress relief is specified per order; it is not implied by the grade designation alone.
If your machine struggles with titanium in general, the blank geometry is often part of the cause. Send the machine model, the planned part geometry and the tooling in use, and the setup is reviewed against the actual machine rather than against a generic recommendation.
Traceability, EN 10204 3.1
Batch Traceability from Mill Heat to Shipped Blank
Every disc batch is released against a heat or lot reference carried through incoming inspection, machining, vacuum annealing, final dimensional verification and shipment. Marking, document set, protective packing and delivery basis are checked before release, so the certificate, the geometry and the blank can be reconciled against one order basis.
Heat or lot level traceability from incoming bar or plate
EN 10204 3.1 standard, 3.2 with independent witness on request
Marking, packing list and release documents agreed in advance
A certificate name alone does not define the complete acceptance package. State the traceability level, the inspection characteristics, the third-party involvement and the release documents required by the order.
🧾
EN 10204 3.1 mill test report
Every batch ships with a 3.1 MTR declaring the chemistry and mechanical results of the actual heat. Type 3.2 with an independent inspection witness can be arranged where the customer process requires it.
🔗
Heat-to-blank batch traceability
The incoming bar or plate heat identity is carried through machining, inspection and shipment, so the certificate, the geometry and the blank can be reconciled against one order basis.
🏥
ISO 13485 partner arrangement
Disc blanks are manufactured under a contract manufacturing partner holding ISO 13485:2016, authorised to supply BOZE. The certificate scope and the authorisation letter are available on request. The certificate is issued to the partner, not to Baoji Boze Metal Products Co., Ltd.
🔬
Material, not device, conformity
The blank conforms to the material standards ASTM F67 or ASTM F136 and the ISO 5832-2 / 5832-3 equivalents. A material conformity statement is not a finished medical device approval, and device-level certification remains with the dental laboratory or manufacturer.
Before You Order
Confirm Four Controls Before You Buy
Use this short check before comparing quotations. It separates the four questions buyers most often conflate: interface, standard revision, documentation set, and the scope of the material claim.
Verify the interface
Confirm the blank diameter and locating feature against the actual machine model before the order, not against the reseller listing.
Name the standard revision
ASTM F67 and ASTM F136 have revisions, and the MTR must declare the revision in force for the order. State it explicitly.
Confirm the documentation set
Agree the mill certification level, the inspection characteristics and the third-party witness requirement in advance of manufacture.
Separate material from device claims
Ask for material conformity to ASTM F67 / F136 and the material test report. A QMS certificate is a different document answering a different question.
Supply Chain
Sourcing, Machining, Vacuum Processing and Release
Four supply-side controls that determine whether a specific blank order can be accepted, and on what schedule.
🏭
Material sourcing
CP Grade 1-4 and Grade 23 ELI bar and plate are sourced against the specification named on the order, with the mill identity recorded on the MTR.
⚙️
Precision machining
Blanks are turned and milled to the interface geometry and tolerance, with the register step and locating features produced in the same setup as the face and OD.
🌡️
Vacuum processing
Vacuum annealing relieves residual stress so the disc stays dimensionally stable under the load of a five-axis finishing cut.
🔍
Inspection and release
CMM dimensional verification, hardness and surface condition checks against the order, with the document pack released against the same order reference.
Product Form Selection
Disc Blank, Bar and Rod, or Plate and Sheet
Three product forms answer the same underlying need. The controlled purchase description is the exact form, the exact grade, the exact interface and the exact document scope.
Medical Device QMS — Authorized Contract Manufacturing
MTR Traceability
Material Test Reports with Every Shipment
Quality
Uncompromising Quality Control
Every titanium part is verified, documented, and traceable before it leaves our Baoji facility — because in
aerospace and medical manufacturing, documentation is just as critical as the cut itself.
CMM Inspection
Coordinate Measuring Machines verify every critical dimension against GD&T to ±0.005 mm before any part ships.
FAIR Reports
Project-specific First Article Inspection Reports delivered for every new part number — engineered for fast PPAP / FAIR approval.
Material Traceability
Material Test Reports (mill certificates) accompany every shipment for full source-to-component traceability.
Frequently Asked Questions
Dental Titanium Disc Blank FAQ
Eight buyer questions on grades, machine interfaces, certification scope, mechanical properties, interstitial control, inspection, milling practice, and order quantities.
Q1What material grades are available for dental titanium discs?
Commercially pure Grade 1, 2, 3 and 4 to ASTM F67 and ISO 5832-2, and Grade 23 Ti-6Al-4V ELI to ASTM F136 and ISO 5832-3. Grade 2 is the most widely specified CP grade for crowns and abutments; Grade 23 ELI is specified for implant abutments and screw-retained frameworks where fatigue-critical behaviour matters. The grade, the standard revision and the mechanical results must all be declared on the mill test report for the order.
Q2What diameter blanks fit my CAD/CAM machine?
Three published interfaces are supplied: the open system at Ø98 mm nominal with a central register step, the Zirkonzahn system at Ø95 mm, and the Amann Girrbach system at Ø89 mm with a peripheral side notch. The Ø98 mm open blank fits Roland, VHF, Imes-Icore and Yenadent chucks. These geometries are not interchangeable, and the brand names are referenced only to identify the published chuck geometry — they imply no affiliation or endorsement.
Q3Is your ISO 13485 certification available for dental titanium discs?
The manufacturing is performed by our authorised contract manufacturing partner, which holds ISO 13485:2016. The certificate is issued to that partner, not to Baoji Boze Metal Products Co., Ltd., and the authorisation letter and the certificate scope are available on request. We do not claim to hold the certificate ourselves. For the product itself, material conformity is carried by ASTM F67 or ASTM F136, the ISO 5832-2 or 5832-3 equivalent, and the EN 10204 3.1 mill test report.
Q4What mechanical properties should I expect from a Grade 23 ELI dental disc?
For Grade 23 Ti-6Al-4V ELI in the annealed condition: minimum tensile strength 860 MPa (125 ksi), minimum yield strength at 0.2% offset 795 MPa (115 ksi), minimum elongation 10%, and fracture toughness KIC of 70-85 MPa√m. For commercially pure Grade 2 the corresponding minima are 345 MPa tensile, 275 MPa yield and 20% elongation. These are typical annealed-condition minima; the actual values for your order are the ones declared on the mill test report.
Q5How is oxygen and other interstitial content controlled?
Oxygen is the dominant interstitial in titanium and the primary determinant of the ductility-versus-strength trade-off. Grade 23 ELI is held to a maximum oxygen content of 0.13%, and nitrogen, carbon and hydrogen are controlled to the same low-interstitial envelope. On commercially pure Grade 2, mill-stock supply is typically at a 0.13% maximum. Every heat is reported against the specification on the mill test report, and it is that declared result which governs acceptance rather than the grade name alone.
Q6What inspection is performed on the blanks before shipment?
Dimensional verification by CMM covering outside diameter, thickness, register step height and concentricity, plus hardness reported per ASTM E18 and surface condition verified against the order. Dye penetrant inspection to ASTM E1417 is applied where surface-defect acceptance is specified. Ultrasonic testing to ASTM E2375 is not applicable to a solid disc blank, because that standard addresses tubular product and a solid blank has no internal geometry to interrogate. Any additional inspection scope is agreed on the order.
Q7How should titanium disc blanks be milled to avoid tool wear?
Three principles govern the setup. Keep the cutting speed well below the equivalent steel value, because titanium moves heat into the workpiece rather than into the chip. Keep the feed per tooth stable and relatively high for the tool diameter rather than reducing it, since a light dwelling feed work-hardens CP titanium and accelerates edge breakdown. Deliver coolant at the cutting edge under pressure, either flood or through-spindle, because starved cooling is the most frequent cause of titanium edge burn. Variable-helix tooling and a rigid setup complete the strategy. Absolute surface speeds and feeds are confirmed against your machine rigidity, holder, tool diameter and blank grade during the setup review rather than taken from a generic table.
Q8What are the minimum order quantity and lead time?
Both are quotation-specific. They depend on the grade, the interface geometry, the blank dimensions, the quantity, the current bar and plate availability and the machining capacity of the order. Sample quantities are generally reviewable for engineering evaluation. Treat the confirmed values in the quotation as the operative figures rather than treating any published lead time as a commitment.
Dental Titanium Disc Inquiry
Send the Order Basis for Supply Review
Share the machine model, the interface, the grade and specification revision, the blank geometry, the quantity and the evidence requirements. The machine model and the grade are the two inputs that determine whether the order can be accepted without a clarification round.
Machine model and blank interface
Grade and standard revision
Blank thickness and geometry
Quantity and documentation scope
Continue Your Specification Path
Related Stock, Capabilities, and RFQ
Five next-step anchors: parent hub, the upstream bar stock, the grade library, traceability evidence, and the quotation route.
Founded in 2011 in Baoji's Titanium Valley, BOZE Metal is dedicated exclusively to titanium — from raw material to precision engineering.
AS9100D, & ISO 9001 certified
with 500+ clients across Aerospace, Medical & Motorsport industries, we deliver end-to-end precision titanium CNC machining with full material traceability from source to component.
Boze Titanium Manufacturing Center is operated by Baoji Boze Metal Products Co., Ltd.
Certifications verified by
ASTM International titanium material specifications (B265, B348, B381, F136, F2924) with full EN 10204 3.1 mill test certification. Our quality system is operated under AS9100D by Baoji Boze Metal Products Co., Ltd.
This website uses cookies to improve your browsing experience, analyze site traffic, and deliver personalized advertising. By clicking 'Accept All', you consent to the use of all cookies.